2026265-075C Solar 8000M/i patient monitor 5-21
Troubleshooting: Error messages
“Real Time Clock FAILED - will not start.”
“WARNING: The real time clock chip is not running. Main Code
cannot be loaded until this chip is started. Attempting to start
real time clock...”
Followed by either
“The real time clock was started. Select Start Patient Monitoring
to load and execute Main Code.”
Or
“ERROR: Unable to start the real time clock.”
If problem persists and error message displays, replace
processor pcb.
“EC1” (displayed to the left of the ECG parameter block) Perform all of the maintenance procedures listed in the
“Manufacturer Recommendations” section of Chapter 4,
Maintenance.
Service the PDM - Error Code The Patient Data Module (PDM) has encountered a
software, hardware or both a software and hardware error.
Remove the PDM from the patient monitoring system
immediately and send to service for repair. See the Patient
Data Module service manual for troubleshooting information.
Message Possible reason and solution