A-14 Solar 8000M/i patient monitor 2026265-075C
Appendix A – Technical specifications: Tram modules and Solar parameter functionality
Analog output: Selectable saturation 0 to 100% equivalent 0 to 1V
Alarm limit range: SpO2: 0 to 105%; PPR: 0 to 350 beats per minute.
Messages sent to host:
GE: Pulse search, Poor signal quality detected, Probe is off patient, Probe or module malfunction
Masimo: Probe or module malfunction, Artifact detected, Probe is off patient, Low quality, Pulse search
Nellcor: Probe is off patient, Low quality, Pulse search
Cardiac output
Method: Thermal dilution
Cardiac output range: 0.2 to 15 liters per minute
Blood temperature range: 30°C to 42°C (86°F to 107°F)
Injectate temperature range: 0°C to 30°C (32°F to 86°F)
Waveform display frequency response: 0 to 10 Hz (-3 dB)
Output parameters: Cardiac output, blood temperature, injectate temperature, real-time cardiac output
waveform
Cardiac output review: Accept / reject individual measurements and store average
Catheter sizes: 5, 6, 7, 7.5, 8, French
Injectate volume selection: 3, 5, 10 cc
Environmental specifications (all modules)
Power requirements: 16.75 VDC
Power consumption: 10 Watts (module alone)
Cooling: Forced air
Heat dissipation: Tram 451, 451M, 451N, 851, 851M, 851N: 34 Btu/hr
All others: 31 Btu/hr
Operating temperature: 10°C to 35°C (50°F to 95°F)
Operating humidity: 40% to 95% (noncondensing)
Storage temperature: -10°C to 50°C (14°F to 122°F)
Storage humidity: 0% to 95% (noncondensing)
Physical specifications (Tram modules)
Height: 8.1 cm (3.2 in)
Width: 11.4 cm (4.5 in)
Depth: 29.5 cm (11.6 in)
Weight: 2.1 kg (4.7 lb)