2026265-075C Solar 8000M/i patient monitor 4-25
Maintenance and checkout: Checkout procedure
Cardiac output functions
1. Connect the cardiac output cable adapter to the Temp/CO connector
of the acquisition device.
2. Connect the cardiac output adapter to the cardiac output cable
adapter and to the simulator.
3. At the patient monitor, go to MENUS CO.
4. Set up the monitor:
Set AUTO MODE: to ON
Set INJECT TEMP: to BATH
Set SIZE: to 7
Set INJECT VOL: to 10CC
Set COMPUTATIONAL CONSTANT: to 0.540
5. Select cardiac output menu corresponding to the injectate
temperature. (CO)
6. Set the adapter and simulator to . Verify that the IT value on the
lower left of the display matches the input.
When computing is complete, the CO reading on the patient monitor
should read approximately ±5% of simulated value.
7. Repeat this test with the adapter and simulator set to 24° and the
COMPUTATIONAL CONSTANT: set to 0.595.
The computed output should be ±5% of simulated value.
SpO
2
functions (Masimo)
1. Connect the Masimo SpO2 tester cable to the SpO2 sensor adapter
cable.
2. Connect the SpO2 adapter cable to the SpO2 connector on the
acquisition device.
3. Turn the patient monitor on.
4. Verify the following are displayed at the patient monitor: (It might
be necessary to turn the SpO
2
parameter on.)
A waveform with an SpO
2
label.
An SpO
2
% reading between 78- 84%.
A PRR reading between 60 and 62 beats per minute.
5. Disconnect the simulator cable from the acquisition device.
SpO
2
functions (Nellcor OxiMax)
1. Verify that 2 AA alkaline batteries are installed in the NELLCOR
SRC-MAX Pulse Oximetry functional tester.
2. Connect the SRC-MAX to the SpO
2
connector on the module.
3. On the SRC-MAX, verify that the IR and RED LED indicators are
both lit.