4-32 Solar 8000M/i patient monitor 2026265-075C
Maintenance and checkout: Checkout procedure
NOTE
The Marker Out amplitude and pulse width are configured at 10v
and 10ms respectively from the factory.
Arterial BP
024A
Signal pin: 2 Red wire
Ground pin: 6 Green wire
Time/division: 0.2s
Volts/division: 0.2v
Marker Out (Frequency)
025A
Signal pin: 9 Gray wire
Ground pin: 4 Yellow wire
Time/division: 10ms
Volts/division: 0.5v
Marker Out (Pulse Width)
026A
Signal pin: 9 Gray wire
Ground pin: 4 Yellow wire
Time/division: 5ms
Volts/division: 1v