Technical Specifications
2001589-312 Revision B TONOPORT VI 29
9 Technical Specifications
Measuring Range
systolic pressure: 60 to 260 mmHg
(8.0 to 34.6 kPa)
diastolic pressure: 40 to 220 mmHg
(5.3 to 29.3 kPa)
mean pressure: 50 to 250 mmHg
(6.7 to 33.3 kPa)
pulse rate (HR): 35 to 240 min
-1
Measurement Accuracy
(determined in a clinical study)
systematic measurement deviation
for deflation measurement method:
0.2 mmHg (systolic)
0.1 mmHg (diastolic)
empirical standard deviation
for deflation measurement method:
2.8 mmHg (systolic)
2.9 mmHg (diastolic)
systematic measurement deviation
for inflation measurement method:
-0.1 mmHg (systolic)
0.5 mmHg (diastolic)
empirical standard deviation
for inflation measurement method:
3.6 mmHg (systolic)
2.4 mmHg (diastolic)
Measurement Capacity
up to 400 blood pressure measurements
Interfaces
USB (1.1 or 2.0)
RS 232 (9600 Bd / 8N1)
Battery
2 AA size rechargeable NiMH batteries, 1.2 V,
>1500 mAh or
2 AA size alkaline batteries
Battery Charge Time
2 to 3 hours
Maximum Cuff Pressure
300 mmHg
Measurement Method
oscillometric, selectable measurement method:
deflation measurement method or inflation
measurement method
Battery Charger
protection class II, IP20
100 to 240 VAC 50/60 Hz, 0.5 A
Ambient Conditions
Operation
temperature between 0 and 55°C
relative humidity between 15 and 93%, no
condensation
atmospheric pressure between 700 hPa and 1060 hPa
altitude (relative to sea level) -400 to 2800 meters
Transport and Storage
temperature between -25 and 70°C
relative humidity between 10 and 93%, no
condensation
atmospheric pressure between 500 hPa and 1060 hPa
altitude (relative to sea level) -400 to 4500 meters
Dimensions and Weight
height 27 mm
width 73 mm
depth 108 mm
weight <190 g, incl. batteries
Protection Class
IP20: TONOPORT VI
IP02: wearable pouch of the TONOPORT VI
IP22: TONOPORT VI in wearable pouch
Expected Service Life
TONOPORT VI: 10 years
cuff: 20,000 cycles of reapplication
Note
The device needs 30 min to get ready for its intended
use and reach the operation conditions form the
minimum storage temperature and the maximum
storage temperature, if the room temperature is
20°C.

Related product manuals