Chapter 2. Installation
TransPort
®
PT900 User’s Manual 25
2.5.6b Odd-Traverse Installations (cont.)
6. Set the yoke axial position as follows:
• Spacing >305 mm/12 in. (435 mm/17.13 in. is used as an example):
Measure the required 435 mm/17.13 in. distance from the edge of the
fixed clamp to the marked line on the yoke (see Figure 29 below).
Figure 29: Yoke Set at 435 mm/17.13 in.
• Spacing <305 mm (120 mm/4.72 in. is used as an example): Measure
the required 120 mm/4.72 in. distance from the edge of the fixed
clamp to the marked line on the yoke (see Figure 30 below).
Figure 30: Yoke Set at 120 mm/4.72 in.
Fixed Clamp at Zero
Marking on Yoke
17.13 in. (435 mm)
Fixed Clamp
Marking
4.72 in.
on Yoke
at Zero