Site Requirements
Venue 50Basic User Manual 2-3
Direction 5446729-100 English Rev. 7
Before the system arrives
NOTICE This medical equipment is approved, in terms of the prevention
of radio wave interference, to be used in hospitals, clinics and
other institutions which are environmentally qualified. The use of
this equipment in an inappropriate environment may cause
some electronic interference to radios and televisions around
the equipment.
Ensure that the following is provided for the new system:
A separate power outlet with a 6 amp circuit breaker for
220-240 VAC or a 10 amp circuit breaker for 100-120 VAC.
NOTE: This is for the Docking Station/Cart.
Take precautions to ensure that the console is protected
from electromagnetic interference.
Precautions include:
Operate the console at least 15 feet away from motors,
typewriters, elevators, and other sources of strong
electromagnetic radiation.
Operation in an enclosed area (wood, plaster or
concrete walls, floors and ceilings) helps prevent
electromagnetic interference.
Special shielding may be required if the console is to be
operated in the vicinity of radio broadcast equipment.