Powering the System
Venue 50Basic User Manual 2-27
Direction 5446729-100 English Rev. 7
Powering the System
Connecting and Using the System
To connect the Docking Station/Cart to the electrical supply:
1. Ensure that the wall outlet is of the appropriate type.
2. Push the power plug securely into the wall outlet.
CAUTION
Use caution to ensure that the power cable does not
disconnect during system use.
If the system is accidentally unplugged, data may be lost.
WARNING
DO NOT use the Venue 50 on plastic foam, paper or similar
type surfaces. The system could overheat and slow down.
Ensure that the Venue 50 is on a sturdy, heat resistant surface.
WARNING
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet. See ‘Before the system arrives’
on page 2-3 for more information.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than
specified. Never use additional multiple portable socket-outlets,
an extension cord or an adapter plug.
To help assure grounding reliability, connect to a “hospital
grade” or “hospital only” grounded power outlet.