Preparing the System for Use
2-38 Venue 50Basic User Manual
Direction 5446729-100 English
Rev. 7
Connecting the Probe (continued)
5. Carefully position the probe cord so it is free to move and is
not resting on the floor.
6. When the probe is connected, it is automatically activated.
Cable Handling
Take the following precautions with probe cables:
If the system in on the docking cart, please keep the probe
cable free from wheels and use the cable hooks on the
docking cart.
DO NOT bend the cable acutely.
Avoid crossing cables between probes.
Deactivating the Probe
When deactivating the probe, the probe is automatically placed
in standby mode.
To deactivate a probe:
1. Ensure the Venue 50 is in freeze mode or powered off.
2. Gently wipe the excess gel from the face of the probe.
3. Carefully slide the probe toward the probe holder on the
docking station. Ensure that the probe is placed gently in the
probe holder.
CAUTION
Fault conditions can result in electric shock hazard. Do not
touch the surface of probe connectors which are exposed
when the probe is removed. Do not touch the patient when
connecting or disconnecting a probe.