General procedures
Venue 50 – Basic Service Manual 4-7
5447566-100 English Rev. 7
Power shut down
NOTE: After truning off a system, wait at least ten seconds before
turning it on again. The system may not be able to boot if power
is recycled too quickly.
Follow these steps to shut down the system:
1. Slightly press the Power On/Off switch once.
2. The System-Exit window is displayed.
3. Select OK.
4. The shutdown process takes several seconds and the
power off sequence is complete when the power status LED
is turned off.
5. Disconnect the probes. Clean or disinfect all probes as
necessary. Store them in their shipping cases to avoid
damage.
Adjusting the Display Monitor
See ‘Adjust Brightness’ on page 6-3 for more information.
CAUTION
DO NOT unplug and/or transport the unit until after the power
off sequence has been completed. Failure to do so may result
in corrupted patient files.