Introduction
1-20 Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Electrical safety
Safe practices
Follow these guidelines to minimize shock hazards whenever
you are using the Ultrasound system:
To minimize shock hazard, the equipment chassis must be
connected to an electrical ground.
The Ultrasound system is equipped with a three-conductor
AC power cable. This must be plugged into an approved
electrical outlet with safety ground.
The power outlet used for this equipment should not be
shared with other types of equipment.
Both the system power cable and the power connector must
meet international electrical standards
Probes
Follow these guidelines before connecting a probe to the
Ultrasound system:
Inspect the probe prior to each use for damage or
degradation to the:
• housing
cable strain relief
• lens
• seal
• connector pins
• locking mechanism
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any
liquid.
Peripherals
Refer to the Patient Safety Environment section of the User’s
Manual for peripheral isolation information.
WARNING
Connecting a Venue 50 to the wrong voltage level will most
likely destroy it.