D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-5
PRELIMINARY
8-2-3 Back Cover Replacement Procedure
8-2-3-1 Tools
Phillips screwdriver of appropriate size.
8-2-3-2 Time Required
5 min
8-2-3-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-3-4 Back Cover Removal Procedure
1) Remove batteries.
2) Remove Probe Holder Rubber by gently pulling it upwards starting from the middle.
Refer to Table 9-16 on page 9-17.
• Batteries Removal Procedure
Figure 8-73 Probe Holder Rubber Removal