D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
8-12 Chapter 8 - Replacement Procedures
PRELIMINARY
5) Slip the broken probe lever out;
8-2-5-5 Probe Lever Installation Procedure
6) Attach the black rubber seal to the lever. Then slip the lever into the axle.
Note: The lever has a longer (marked in Blue) and shorter arc. Ensure the lever is inserted with
its long arc as shown in the picture below:
7.) Slip the black plastic cap to secure the lever on the axle.
8.) Attach 2 screws holding the probe lever cap.
9.) Test the functionality of the probe locking before proceeding.
10) Install C-PSB.
11.)Install back cover.
Figure 8-83 Slip Probe Lever Out and remove rubber seal
Figure 8-84 Insert Black Rubber Seal and Place Lever on the Axle
C-PSB Installation Procedure
Back Cover Installation Procedure

Table of Contents