D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-55
PRELIMINARY
5.) Disconnect 2 screws from the ECG bracket.
6.) Remove the ECG assembly from the cart; Then pull the ECG out of its bracket.
8-2-22-5
ECG Installation Procedure
1) Install the ECG in its bracket.
2.) Install the ECG assembly on the cart, and re-connect two screws.(Note: the ECG may be installed
on either side of the cart).
3.) Connect the USB cable to the ECG.
Figure 8-135 Remove ECG Bracket Screws
Figure 8-136 Remove ECG