System Warning/Error and Logs
Versana Balance – Basic Service Manual System view 7-9
5808768-100 English Rev.9
7-4 System Warning/Error and Logs
7-4-1 Temperature warning
Figure 7-8. Temperature Warning
Cause: CPU temperature exceeds threshold (95 degrees)
What to do: shut down the system and clean the filter.
7-4-2 Temperature exceeds threshold
Figure 7-9. Temperature exceeds threshold
Cause: CPU temperature exceeds threshold (100 degrees)
and therefore, system must be shutdown immediately.
What to do: shut down the system and clean the filter.