Loading the software
Versana Balance – Basic Service Manual System view 8-9
5808768-100 English Rev.9
8-4 Loading the software
8-4-1 Purpose of this section
This section describes how to reinstall software on Versana
Balance.
8-4-2 Customer provided prerequisite
• Formatted and labelled media for Images storage.
• Formatted and labelled media for Patient Archive and
Presets (User Defined Settings).
• Password for the user ADM.
8-4-3 Data Management - moving all images
• Move the images to a remote server or to removable media.
For instructions, please see “Disk management” in the Basic
User Manual/User Guide.
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from Versana Balance before doing backup of the
Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.