EasyManua.ls Logo

GE Versana Balance - Page 417

GE Versana Balance
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical safety tests
Versana Balance – Basic Service Manual System view 10-27
5808768-100 English Rev.9
10-7-5-2 Data Sheet for enclosure Source Leakage Current
The test passes when all readings measure less than the value
shown in
Table 10-13 on page 10-22. Record all data on the PM
Inspection Certificate.
Table 10-17: Typical Data Sheet for enclosure Source Leakage Current
Unit
Power
Tester
Polarity
Switch
Tester
Neutral
or
Ground
Switch
Test 1
Speaker
Cover
Test 2 Real
Panel Metal
Parts
Optional
Test 3
Optional
Test 4
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED

Table of Contents

Related product manuals