GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 3 Installation 3-51
3-9-3 Setting Up for Connectivity
NOTE: Four passwords are required to complete the setup of the connectivity options. Ensure that all
these passwords are installed on the Vivid 3N Pro/Expert ultrasound unit before you begin the
configuration procedure.
1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
2) Trackball to the System Options tab and press Select. The System Options tab is displayed, as
shown below:
Figure 3-33 System Options Tab
3) Verify that the following options are installed, as required:
• DICOM Media Creator is required to support DICOM storage to MOD or CD-R media.
• DICOM Network Connectivity is required to support DICOM storage over the network, for
example, to Proslov.
• EchoNET Connectivity is required to support an EchoPAC direct connection.
• Excel Export is required to support the ability to export measurements and reports to Microsoft
Excel files.
4.) Enable the DICOM printer option, as required (under the Services sub-tab) - see Figure 3-34.