IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
3-56 Section 3-9 - Connectivity Setup
3-9-5 Setting Up for Communication with a Prosolv Workstation
The Vivid 3N Pro/Expert ultrasound unit can connect to a Prosolv DICOM viewer through a DICOM
server. The unit cannot connect directly to the viewer.
1) Setup the network connection, see 3-9-4 "Setting Up the Network Connection" on page 3-54.
2) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
3) Trackball to the Connectivity tab and press Select. The Connectivity tab is displayed, with the
Views subsidiary tab selected by default, as shown in Figure 3-36 "Connectivity Tab" on page 3-54.
4) Trackball to the SERVICE button at the bottom of the Views tab, and press Select. The password
request dialog box, shown in Figure 3-37 "Password Request Dialog Box" on page 3-54, is
displayed.
5) Enter the password geulsservice, and then trackball to the OK button and press Select.
6) Trackball to the Services tab and press Select. The Services tab is displayed, as shown below:
Figure 3-39 Services Tab
7) Select DICOM Server from the Name dropdown list. The Services tab is updated, as shown below
Figure 3-40 Services for DICOM Server