GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
5-18 Section 5-4 - Front End
5-4-5 MUX Board
5-4-5-1 General Description
The MUX Board has 64 identical receive channels, and performs the following functions:
Receiving the channels and connecting them to the BF Boards.
Expanding the low/high channels in CW mode.
TEE motion control.
Monitoring temperature and voltage measurements via the I²C.
FE Bus switch. To reduce the noise in the FE cards, the MUX switches off the FE Bus to the FB
during the data acquisition in CW mode.
Min/max AC input monitoring, every ten seconds.
Figure 5-11 MUX Board Diagram
Input Channels
Multiplex
TEE
Control
Monitor
Temp. Voltages
Fan Operation
FE Bus Switch
FE
Bus
FB
IC
Interface
2
FB
FB
BF

Table of Contents