EasyManua.ls Logo

GE VIVID 3 BT00 - RFI Leds

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 6 Service Adjustments 6-5
6-3-1 RFI LEDs
Table 6-3 RFI LEDs
LED LED color Description Normal State
LD505
Red
Demod Error
LD504
Green
Lit in Scanning Mode Slow tick in idle
LD503
Yellow
Transmit data to BEP During Processing
LD502
Red
Not Used
LD501
Yellow
Not Used
LD500
Green
Not Used
LD12
Green
5.0 V status
LD22
Green
2.0 V status
LD11
Green
1.4 V status
LD10
Green
1.5 V status
LD9
Green
3.3 V status
LD8
Green
5.0 V status
LD7
Green
2.5 V status

Table of Contents