GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
6-6 Section 6-3 - Front End Voltages and Signal Indicators
6-3-2 Image Port (IMP) LEDs
6-3-3 Front End Controller (FEC) LEDs
6-3-4 RF and Tissue Processor (RFT)
Table 6-4 Image Port LEDs
LED Indicates When Lit Comments
Red Board Failure Normally ON
Yellow
Yellow VME Master Lit during scanning and data transfer
Yellow VME Slave
Green Board OK Normally lit
Red Board Failure Normally OFF
Table 6-5 Front End Controller LEDs
LED Indicates When Lit Comments
Red Board Failure Normally OFF
Green Board OK Normally lit
Yellow Scan Lit during scanning
Yellow Probe Change Lit during probe change
Yellow No Probe Lit if no probe is selected
Yellow HV Over Range
Lit if excessive HV is output,
possibly due to HV surveillance
Table 6-6 RF and Tissue Processor LEDs
LED Indicates When Lit Comments
Red Board Failure Normally OFF
Green Board OK Normally lit
Yellow DSP Runs Normally blinking
Yellow Data Out on Pipeline Lit during data transfer to pipelink
Red Clock Problems Normally OFF
Red Clock Problems Normally OFF
Red Clock Problems Normally OFF

Table of Contents