GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
6-12 Section 6-6 - Monitor Operation
6-6-1-2 15" and 17" Monitor Function Control Options
Table 6-11 Vivid 3N Pro/Expert 15" and 17" Monitor Controls and Functions
Control Function
Adjusts the background Brightness (white level) of the screen.
Adjusts the Contrast level (black level) of the screen.
,A 'Moire' pattern can appear on your screen, looking like a series of
concentric circles or arcs. To eliminate this pattern, use the
horizontal moire adjustments.
A 'Moire' pattern can appear on your screen, looking like a series of
concentric circles or arcs. To eliminate this pattern, use the vertical
moire adjustments.
The Degauss feature will remove color impurities caused by
magnetic fields. Do not use the Degauss feature more than once
within a 30-minute period.
Note: The monitor may buzz momentarily, the image colors may
change and the image will jiggle for a few seconds. These effects are
normal.
Use the Recall feature to reset these monitor settings to their original
levels: Position, Size, Pincushion, Trapezoid, Parallelogram,
Pinbalance, Rotation, Clear Moire.Note: If you have selected "Yes",
all settings listed above will be reset. All other settings will remain the
same.
Color temperature is a measure of the 'warmth' of the image colors.
The available range is between 5000 to 9300K