GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 6 Service Adjustments 6-13
6-6-1-2 15" and 17" Monitor Function Control Options (cont’d)
Table 6-12
Control Function
Adjust the monitor color setting (Red).
Adjust the monitor color setting (Green).
Adjust the monitor color setting (Blue).
When you adjust the Brightness and Contrast after sRGB mode is selected, sRGB mode exits
Follow these instructions to adjust the horizontal position of the monitor’s entire display.
Follow these instructions to adjust the vertical position of the monitor’s entire display.
Follow these instructions to adjust the horizontal size of the monitor’s entire display.
Follow these instructions to adjust the vertical size of the monitor’s entire display.
Adjust the pincushion setting when the sides of the display are bowed in or bowed out.
Adjust the pinbalance setting when the sides of the display are bowed towards the left or right.
Adjust the trapezoid setting when the top or bottom of the display is too large or small.
Adjust the parallelogram setting when the display is leaning left or right.
Adjust the rotation setting when the entire display is tilted left or right.
Adjust the side pin corner correction when the top or bottom of the display is too large or smal