GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-4 Section 7-2 - Diagnostics
7-2-3-1 Diagnostic Menu Options
The Diagnostic menu includes the following options:
Software Version: Displays the Vivid 3N Pro/Expert ultrasound unit software version number and
last upgrade date.
: Displays test results, as described in Section 7-5 "Common Service Interface" on
page 7-88.
: Calibrates the Front End crate, and enables you to perform the display and video
grabbing test, as described in Section 6-8 "Calibration" on page 6-21.
: Tests each board in the Front End crate including the TR power supply, the LVPS
supply and temperatures, as described in Section 7-3 "Performing Front End (FE) Diagnostics" on
page 7-5.
: Tests various back end hardware components, as described in Section 7-4
"Performing Back End Diagnostics on the System" on page 7-71.
: Monitors all low voltages, Tx voltages, temperatures and AC power input.
: Performs a full sequence of Front End tests that do not require user intervention.
Auto Sequence is particularly recommended for detecting problems that are intermittent in nature,
since a test may be paused at any stage and any detected failures reviewed. This is described in
7-3-1-1 "Auto Sequence" on page 7-6.
Description Area: Displays a description of the function performed when each of the buttons is
selected. Trackball the cursor over a button to display a description in the Description area.
Report Button: Displays a current diagnostics report.
Exit Button: Closes the Diagnostic menu.