IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-48 Section 7-3 - Performing Front End (FE) Diagnostics
7-3-8-8 RAM Test
The RAM test enables you to check the RAM that stores the delays for the transmitted ultrasound
signals to the form focal beam.
NOTE: You must reboot the system after performing the RAM test.
1) Access the F.B. tab of the FE Diagnostics dialog box, as described in 7-3-8-1 "Accessing the Front
Board Assembly Options" on page 7-27.
2) Trackball to the RAM Test button and press Select. The RAM test is performed and the results are
displayed in the Comments area, as shown in Figure 7-25 (or Figure 7-26) below::
Figure 7-25 F.B. Tab - FE Diagnostics Dialog Box - RAM Test (RFI Configuration)