IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-50 Section 7-3 - Performing Front End (FE) Diagnostics
7-3-9 MUX Diagnostic Tests
NOTE: The Select Socket and Select Supply options are not relevant for this test.
7-3-9-1 Accessing the MUX Diagnostic Test Options
1) Access the FE Diagnostics dialog box, as described in 7-3-1 "Accessing the Front End Diagnostic
Options" on page 7-6.
2.) In the FE Diagnostics dialog box, trackball to the MUX tab and press Select,
OR
3.) Trackball to the MUX button in the Data Flow Maps and press Select. The MUX tab is displayed,
as shown in Figure 7-26 (or Figure 7-27) below: