IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-63
7-3-9-6-2 AC Power Input Test
The AC power input test measures the AC input voltage deviation from the nominal input voltage
(accuracy by percentage).
1) Access the MUX tab of the FB Diagnostics dialog box, as described in 7-3-9-1 "Accessing the MUX
Diagnostic Test Options" on page 7-50.
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in Figure 7-35 "MUX Advanced System Tests Options" on page 7-
61.
3) Trackball to the Monitor button and press Select. The View Power options are displayed, as shown
below:
Figure 7-37 View Power Options