IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-77
7-4-4-1 Manual Extended Keyboard Test (cont’d)
Figure 7-50 KB Panel Tab - Extended Keyboard Test Window
5) Press each of the control panel buttons (except for Select, Menu and the On/Off button), and verify
that the correct buttons are activated on-screen in the KB Panel tab.
6) Push each of the rocker switches, and verify that the correct rockers are activated on-screen in the
KB Panel tab.
7) Rotate each of the rotaries, and verify that the correct rotaries are activated on-screen in the KB
Panel tab.
8) Slide each of the sliders, and verify that only the correct sliders are activated on-screen in the KB
Panel tab.
9) Press the soft keys, and verify that only the correct soft keys are activated on-screen in the KB
Panel tab.
10) Connect the Footswitch to the connector on the front panel of the unit, and press the right, middle
and left of the switch. Verify that the numbers {68}, {69} and {70} accordingly are displayed on the
info bar.
11) When you are finished, trackball to the Exit button and press Select.