GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-100 Section 7-5 - Common Service Interface
7-5-3-2 Log Browser (cont’d)
If required, a log file can be saved using the Save function from this screen (Figure 7-74) (i.e, when you
are connected to the system remotely), or you may save the file to a floppy disk on the system itself.
Only three log files may be reviewed in the field:
•DiagErr
• SystemInfo
•errorlog
For all others, the information contained within the log file would be meaningless since it needs to be
reviewed on an actual system and cannot be analyzed remotely.