GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-100 Section 7-5 - Common Service Interface
7-5-3-2 Log Browser (cont’d)
If required, a log file can be saved using the Save function from this screen (Figure 7-74) (i.e, when you
are connected to the system remotely), or you may save the file to a floppy disk on the system itself.
Only three log files may be reviewed in the field:
•DiagErr
SystemInfo
•errorlog
For all others, the information contained within the log file would be meaningless since it needs to be
reviewed on an actual system and cannot be analyzed remotely.