IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-112 Section 7-5 - Common Service Interface
7-5-4-4 Front End Diagnostics (cont’d)
For information on performing a specific Interactive test, refer to the appropriate section, as follows:
Transmit Test - see 7-3-8-2 "Transmit Test" on page 7-29.
FLA Jumper Test - see 7-3-8-3 "FLA Jumper Switch" on page 7-31
TR Switch Test - see 7-3-8-5 "TR Switch Test" on page 7-41.
NOTE: On the TR Switch Test, Socket 1 and Socket 2 relate to which socket the probe is plugged into.
Select the appropriate Pulser.
Pencil Probe Test - see 7-3-8-6 "Pencil Probe Test" on page 7-43.
7-5-5 Image Quality Page
When the Image Quality page is selected, as shown below, you can verify and calibrate image quality.
Figure 7-84 Image Quality Page
NOTE: This page is not populated in this version.
7-5-6 Calibration Page
When the Calibration page is selected, as shown below, you can perform different calibration
procedures, as described in Chapter 6, Service Adjustments.
Figure 7-85 Calibration Page
NOTE: This page is not populated in this version.