IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-16 Section 8-2 - Cover Replacement Procedures
8-2-9 Speaker Cover Replacement Procedure
8-2-9-1 Tools
Use the appropriate Phillips-type screw drivers, as indicated in the speakers cover replacement
procedures.
8-2-9-2 Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
8-2-9-3 Speaker Cover Removal Procedure
1) Remove the bottom keyboard cover, as described in 8-2-8-3 "Bottom Keyboard Cover Removal
Procedure" on page 8-14.
2) Remove the two screws located beneath the keyboard on either side, and the fastening bracket in
the middle, that secure the speaker cover.
Figure 8-17 Speaker Cover Screws
3) Remove the speaker cover.
8-2-9-4 Speaker Cover Installation Procedure
1) Return the speaker cover to its original position.
2) Fasten the cover with two screws and bracket in the middle, beneath the keyboard. Before
tightening the fastening screws, adjust the speaker cover to fit properly and symmetrically.
3) Install the bottom keyboard cover, as described in 8-2-8-4 "Bottom Keyboard Cover Installation
Procedure" on page 8-15.