GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-39
8-3-5 Keypad Replacement Procedure
8-3-5-1 Tools
Use a small flat screw driver, as indicated in the keypad replacement procedure.
8-3-5-2 Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
8-3-5-3 Keypad Removal Procedure
1) Carefully place a small flat screw driver between the keypad you wish to remove and its neighboring
keypad - see Figure 8-39 below.
2) Use the screw driver to gently lever off the defective keypad, until it is completely loosened from its
base.
3.) Lift off and remove the defective keypad.
8-3-5-4 Keypad Installation Procedure
1) Carefully place a new keypad into position on the keyboard, taking care to align the two notches on
the base of the pad with those on the securing base plate.
NOTE: The notches must be vertically aligned with the base, otherwise the keypad will not sit at the correct
height on the keyboard.
2) Gently push the keypad down until it snaps into position.
Figure 8-39 Keypad Replacement