IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-61
8-5-1-4 Back End Processor Installation Procedure (cont’d)
Figure 8-56 System Configuration - Technical Support Tab
26.)Select the appropriate settings, as follows:
* Ex Keyboard - Revision 4
* Sound Board - Ver. 3
* Number of Fans - Set to 7 for BT00 systems; Set to 5 for BT01 systems
* UPS - Change the UPS setting to Used or Not Used, as appropriate.
If the Plug & Scan option was purchased, the UPS setting can be
selected as Used. If this option not purchased, select Not Used.
* VIC - Set to PC VIC
* Nominal Voltage - Set to the appropriate local voltage setting.
27) Perform a complete functional check of the system, as described in Chapter 4 - - Functional
Checks.
28) Install the Vivid 3N rear, left and right covers - refer to the 8-2-2-4 "Side Covers Installation
Procedure" on page 8-4 and 8-2-4-4 "Rear Cover Installation Procedure" on page 8-7.