GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-87
8-5-13 PC-VIC Replacement Procedure
NOTE: This procedure is performed for systems with a BEP2 (Pentium 4-based BEP, newer type).
8-5-13-1 Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the PC-VIC
replacement procedure.
8-5-13-2 Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
8-5-13-3 PC-VIC Removal Procedure
1) Remove the left cover, as described 8-2-2-3 "Side Covers Removal Procedure" on page 8-4.
2) Remove the BEP side cover as described in the 8-5-3-3 "BEP2 - Side Cover Removal Procedure"
on page 8-64.
3.) Remove the screw that holds the PC-VIC in position (as shown in Figure 8-74 below).
4) Carefully slide the PC-VIC out from the unit.
5.) Disconnect all the cables at the rear of the PC-VIC module and the hard disk.
6.) Remove the four screws (two on each side) from the side of the PC-VIC module holder as shown
in Figure 8-75 "PC-VIC Removal" on page 8-87.
7.) Remove the PC-VIC.
Figure 8-74 PC-VIC
Figure 8-75 PC-VIC Removal