EasyManua.ls Logo

GE VIVID 3 BT00 - Page 473

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-103
8-6-6-5 Up/Down Handle Replacement Procedure (cont’d)
4) Mount the handle mounting brackets onto the chassis using one long and one short screw for each
bracket, as shown in Figure 8-87. Tighten the screws to secure the brackets firmly in position.
Figure 8-86 Fitting Gas Spring Cable to Up/Down Handle
Figure 8-87 Mounting the Brackets onto the Chassis
Securing Nut
Up/Down Handle
Gas Spring Cable
Retaining Bracket
shown in detail
Grooved Hole
(insert cable here)
Gas Spring Cable
Short Screw
Long Screw
Short Screw
Long Screw