GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
2-6 Section 2-2 - Console Requirements
2-2-7 Probe Environmental Requirements
NOTE: System and electronic probes are designed for storage temperatures of -20
o
to +50
o
C. When
exposed to large temperature variations, the probes should be kept at room temperature for
a minimum of 10 hours before use.
Table 2-12 Probe Operation and Storage Temperatures
Electronics PAMPTE
Operation
10 — 40
o
C (50 — 104
o
F) 5 — 42.7
o
C (41 — 109
o
F)
Storage
-20 — 50
o
C (-4 — 122
o
F) -20 — 50
o
C (-4 — 122
o
F)