IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-116 Section 8-8 - Peripherals
Section 8-8
Peripherals
8-8-1 B/W Video Printer Replacement Procedure
NOTE: This section describes the installation and removal procedure for the Sony UP-895MDW B/W Video
Printer.
Figure 8-98 SONY B/W Video Printer
8-8-1-1 Tools
Use the appropriate flat and Phillips-type screw drivers.
8-8-1-2 Manpower
One person - 30 mins.
8-8-1-3 Preparations
1) Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panel cover as shown in 8-2-5-3 "Connector Panels Cover Removal
Procedure" on page 8-8.
3.) Unpack the B/W Video Printer.
4.) Check that the B/W Video Printer AC voltage corresponds to that of your local voltage.
5.) Verify that all the DIP Switches on the rear panel of the B/W Video Printer are in the Down position
as shown in Figure 8-99 "B/W Video Printer, DIP Switches on the Rear Panel" on page 8-117.