GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-129
8-8-3-5 BIOS Setup (cont’d)
5.) Use the Select button (see Figure 8-112 "JVC VCR Front Panel Controls" on page 8-128) to select
MEMORY SW and then press the Set button.
6.) Using the Select and Set buttons, scroll through the various Menus and set the values for each of
the relevant Menus, in accordance with those shown in Table 8-4 below.
NOTE: Work in the same order as that given in Table 8-4 and do not change any settings other than those
listed.
NOTE: In Table 8-4, square brackets [] indicate factory presets. The numbers in parenthesis () are seen on the
LCD display.
7.) When all the above-mentioned settings are correct, press the Menu button, select CLOCK ADJUST
and press the Set button. Using the Select and Set buttons, adjust the settings to the correct time.
When the correct time is visible in the LCD display, press the Menu button, to quit CLOCK ADJUST.
8.) Turn OFF power to the VCR. Remove the AC power cable from the mains power wall outlet and
disconnect the cable from the VCR rear panel.
If an external monitor was used for the BIOS setup procedure, turn OFF power to the external
monitor and disconnect the power cable (connected to the VCR rear panel Monitor Out - Video
socket - refer to Figure 8-111 "JVC VCR - Rear Panel - External Monitor Connection" on page 8-
128).
NOTE: All settings are now stored in the VCR memory and the JVC VCR is ready for installation.
Table 8-4 JVC VCR Menu Switch Settings
Menu No. Name Set Value
100 S-VHS SELECT [AUTO] (1)
101
EDIT SELECT ENHANCED (2)
200
Hi-Fi AUDIO REC.
[ON] (1)
213
A.DUB SELECT [A.DUB] (0)
335
RECORDING MODE
[SP] (0)
338 MENU SW 2
OPEN (1)
352 COUNTER MODE
CTL (1)
500 ON SCREEN DISPLAY OFF (0)
005 AUTO TRACKING
[ON] (1)
204 Hi-Fi OUT AT SEARCH
[MUTE] (0)
212 NORM OUT AT SEARCH
OFF (0)
300 DIRECT EJECT
[ENABLE] (1)
312
MODE AT TAPE END STOP (0
354
BAUDRATE SELECT [9600] (3)
359
16: 9 REC OFF (0)
504
INFORMATION SELECT TIME (0)