GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-141
8-8-4-8 Color Video Printer Settings
NOTE: The settings for the Sony UP-2800P and UP2950MD Color Video Printers are regulated using the Video
Printer Control Panel, as described below.
8-8-4-8-1 Preparations
1.) Turn ON the main circuit breaker located on rear of the Vivid 3N.
2) Turn ON the Vivid 3N by pressing the ON/OFF button on the Control Console.
3.) Open the front door of the Color Video Printer.
4.) Press MENU.
The Preset menu is displayed (see Figure 8-124 below).
There are five setup categories: COLOR (c), LAYOUT (l), PRINTER (p), INPUT (i) and OUTPUT
(o). Each setup category can be selected from the Top menu of each category.
8-8-4-8-2 Color Setup
1.) Select COL using the or button (see Figure 8-124, above). The COLOR ADJUST top menu
is displayed on the Printer Window display (Figure 8-125 below).
2.) In the COLOR ADJUST menu, adjust the required parameters (according to those shown in Table
8-6 "WINDOW SETUP Parameters" on page 8-142) by selecting the parameters one-by-one, using
the or button; for each selected parameter, adjust the value using the or button.
Figure 8-124 Color Video Printer Control Panel and Printer Window Display
Figure 8-125 Color Video Printer Window Display (COLOR ADJUST top menu displayed)