GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-142 Section 8-8 - Peripherals
8-8-4-8-2 Color Setup (cont’d)
3.) Press EXEC on the Printer Control Panel to save the Color settings.
8-8-4-8-3 Layout Setup
1.) Return to the COLOR ADJUST top menu (Figure 8-125 "Color Video Printer Window Display
(COLOR ADJUST top menu displayed)" on page 8-141)), using the button.
2.) Select LAY by pressing the or button. The LAYOUT SETUP top menu is displayed (Figure
8-126 below).
3.) In the LAYOUT SETUP menu, select WINDOW by pressing the or button.
4.) Press the button to display the WINDOW SETUP menu (Figure 8-127 below).
5.) In the WINDOW SETUP menu, adjust the required parameters (in accordance with those shown in
Table 8-6 below) by selecting the parameters one-by-one using the or button; for each
selected parameter, adjust the value using the or button.
:
Table 8-5 COLOR SETUP Parameters
Parameter Setting for PAL Setting for NTSC
DARK -2 -5
LIGHT 0 +2
Figure 8-126 Color Video Printer Window Display (LAYOUT SETUP top menu displayed)
Figure 8-127 Color Video Printer Window Display (WINDOW SETUP top menu displayed)
Table 8-6 WINDOW SETUP Parameters
Parameter Setting for PAL Setting for NTSC
H START -8 -16
V START -8 -8
H WIDTH 816 832
V WIDTH 568 480
LAYOUT SETUP
/c/LAY/p/i/o
WINDOW SETUP