EasyManua.ls Logo

GE VIVID 3 BT00 - Cleaning

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
10-10 Section 10-5 - System Periodic Maintenance
10-5-4 Cleaning
10-5-4-1 General Cleaning
10-5-4-2 Air Filter Cleaning
Regular cleaning of the Vivid 3N scanner’s air filter will ensure that the filter does not become clogged
- this would otherwise cause the system to overheat and reduce system performance and reliability.
NOTE: It is recommended that the air filter cleaning procedure is performed once every 3 months (quarterly).
However, the required frequency for air filter cleaning will vary in accordance with environmental
conditions.
Table 10-8 General Cleaning
Step Item Description
1
External
Panels
On a weekly basis, moisten a soft, non-abrasive folded cloth with a mild, general purpose,
non-abrasive soap and water solution or general purpose disinfectant. Wipe down the top,
front, back and both sides of the unit. Do not spray any liquid directly onto the unit!
2
Monitor On a weekly basis, apply glass cleaner to a soft, non-abrasive folded cloth. Gently wipe the
monitor face.
3
Console Frequent and diligent cleaning of the Vivid 3N Pro/Expert ultrasound unit reduces the risk of
spreading infection from person to person, and also helps to maintain a clean working
environment.
Prior to cleaning, turn OFF the power to the system.
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Be careful not to get the cloth too wet so that moisture does not enter the console.
4
Probe Holder Clean the probe holders with warm water and a damp cloth to remove all traces of gel
(soaking may be required to remove excess gel).
Table 10-9 Air Filter Cleaning
Step Item Description
1
Air Filter
Remove the air filter located at the front of the scanner, above the probe connector sockets.
Clean the air filter as follows: shake it in an area away from the system, wash with a mild
soapy solution, rinse, and air dry.
2
Air Filter
Install the clean air filter.
Refer to Chapter 8 - for air filter installation instructions.