IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 3 Installation 3-41
3-8-7 System Options Tab
1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
2) Trackball to the System Options tab and press Select. The System Options tab is displayed, as
shown below:
Figure 3-24 System Options Tab
NOTE: If an option is not available for the specific ultrasound unit, the option name and Key field is disabled
(grey).
3) Trackball to the Key field next to the name of the option being installed, and use the alphanumeric
keyboard to enter the password supplied by GE.
NOTE: If an incorrect password is entered, if the option has already been correctly installed in the system, it will
be removed.
NOTE: Please note that all passwords should be visible underneath the control console or/and on the green
password (option) certificate. If a new password (option) is required, contact the GE representative to
initiate the appropriate process, so that the relevant password (option) will be installed on the system.
4) Trackball to Apply and press Select. The status of the option changes to Installed.
5) Repeat steps 3 and 4 as required.
6) Trackball to the OK button and press Select.
NOTE: If an option was not initially purchased or installed, it may be acquired and/or installed later. During
option installation, type the relevant password in the appropriate Option window (see Figure 3-24,
above) and click Apply. The Installed status (to the right of the Apply button) will then be visible.