GE MEDICAL SYSTEMS
D
IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Chapter 5 - Components and Function (Theory) 5-9
Figure 5-2: Front End Crate Block Diagram - RFT Configuration
The Front End can be divided into two subsystems, as follows:
1.) The Front End subsystem which includes:
- Front Board
- MUX Board
- BF64 (Beamformer Board)
2.) The Mid Processors subsystem which includes:
- RFI (Radio Frequency Interface Board)
OR
* FEC (Front End Controller Board)
* RFT (Radio Frequency & Tissue Board)
* IMP (Image Port).
Front
Board
TR4 Boards
MUX
Board
64Ch
Beam Former
A/D
Board
RFT
Board
Image Port
Board
FEC
(Front End Controller)
Board
Power Supply
Tx P.S.
L.V P.S
Vingmed manufactured boards
I C bus
2
PCI2IP
I
2
C bus
FE bus
Pipeline
link
VME bus
Front
Board
TR4 Boards
MUX
Board
64Ch
Beam Former
A/D
Board
RFT
Board
Image Port
Board
FEC
(Front End Controller)
Board
Power Supply
Tx P.S.
L.V P.S
Vingmed manufactured boards
I C bus
2
PCI2IP
I
2
C bus
FE bus
Pipeline
link
VME bus
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com