GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-33
3.) Carefully return the keyboard to its original position taking care to align the forked metal aligning
plate with the central nut; tighten the M5 nut.
4) Return the two M4 screws to secure the keyboard matrix to the metal frame (one on each of the left
and right corners - see Figure 8-25 on page 8-26).
5) Reconnect the four keyboard cables to the keyboard (previously disconnected) - see Figure 8-24 on
page 8-26.
6) Return the control console upper cover front, as described in the Control Console Upper Cover
(Front) Installation Procedure on page 8 - 15.
7.) Return the control console lower cover, as described in the Control Console Lower Cover
Installation Procedure on page 8 - 18.
8.) Return the control console lower rear cover, as described in the Control Console Lower Rear Cover
Installation Procedure on page 8 - 19.