GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-53
The dialog box displays the Technical Support Tab as shown in Figure 8-43.
Figure 8-43 System Configuration - Technical Support Tab
37.)Select the appropriate settings, as follows:
* Ex Keyboard - Revision 4
* Sound Board - Ver. 3
* Number of Fans - Set to 7 for BT00 systems; set to 5 for other systems.
* MOD Fan - Set to ON (checkbox activated)
* UPS - Change the UPS setting to Used or Not Used, as appropriate.
If the Plug & Scan option was purchased, the UPS setting can be
selected as Used. If this option not purchased, select Not Used.
* VIC - Set to PC VIC
* Nominal Voltage - Set to the appropriate local voltage setting.
38) Perform a complete functional check of the system, as described in Chapter 4 - Functional Checks.
39) Install the Vivid™ 4 rear, left and right covers - refer to the Side Covers Installation Procedure on
page 8 - 4 and Rear Cover Installation Procedure on page 8 - 7.