DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
4 - 4 Section 4-2 - General Procedures
4-2-2-1 Connect Mains Power to the Scanner (cont’d)
5.) Connect the Power Cable’s female plug to the Power Inlet at the rear of the unit.
6.) Lock the plug in position with the Retaining Clasp.
7.) Verify that the Mains Power Circuit Breaker is in OFF position, if not, switch it OFF.
8.) Connect the Power Cable’s other end (male plug) to a hospital grade mains power outlet with the
proper rated voltage, and the unit is ready for Power ON/Bootup.
.
Figure 4-1 Female power plug with retaining clasp off (to the left) and on (to the right)
Figure 4-2 Mains Power Circuit Breaker is in OFF position
Retaining
Clasp off
Retaining
Clasp on
OFF Position