GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 69
5-3-21-22 Test Points
Table 5-58 RFI - Test Points
TEST POINT
# DESCRIPTION COMMENTS
1
EMIFB CLOCK
133 MHz
2
Doppler Probe Sense Voltage
3
Tempsense Output
130.8 mV/
o
C
4
MAX158 INT
NOT USED
5
MAX158 RDY
NOT USED
6
-5 V
7
CLK40
40 MHz
8
N/A
N/A
9
2.5 V voltage
Locally generated from +3.3 V
10
DSP CLOCK
50 MHz
11
15 V MONITOR voltage
12
EMIFA CLOCK
133 MHz
13
DSP_BSOE3_L signal
14
1.4 V voltage
Locally generated from +3.3 V
15
OPTIONAL EMIFB CLOCK
16
N/A
17
N/A
18
N/A
19
DSP CLOCK EMIFB OUT
20
1.5 V voltage
Locally generated from +3.3 V
21
PCI P_LOCK_L signal
22
-15 V voltage
23
+15 V voltage
24
+ 6 V voltage
25
+3.3 V
26
+10 V
27
1.4 V voltage
28
MAX158 INT
NOT USED
29
MAX158 RDY
NOT USED
30
Test Signal -V
Set Amplification
31
Test Signal +V
32
1.5 V voltage