DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-91
15.)Check that the sub-test passed and that result values are within the pre-defined Upper and Lower
limits, as shown in Figure 7-62.
16.)Select the fifth report (Tx Voltages); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-63.
Figure 7-62 View Test Log - Full System Monitoring Test - Battery Parameters