IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
7-54 Section 7-4 - Performing Front End (FE) Diagnostics
The signal should be within the limits marked in the white area (the dotted line represents the
Average), as shown in Figure 7-28.
8) When done, trackball to the Close button and press Set to close the View Report window.
9.) Select the second report (the report is highlighted in the Reports list, and the corresponding results
are shown in the Report window below, as shown in Figure 7-29).
10.)Use the tools provided to scroll through the channels and view the results, as previously described.
11) When done, trackball to the Close button and press Set to close the View Report window.
12.)Select the third report (the report is highlighted in the Reports list, and the corresponding results are
shown in the Report window below, as shown in Figure 7-30).
Figure 7-29 View Test Log - Second Harmonic